| Catalog Number |
323.061 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Misassembled During Installation (4049)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.There was patient involvement associated with the reported event since the event occurred.Date of event is unknown.Complainant device is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.A review of the device history records has been requested and is currently pending completion.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date one (1) 2.0mm threaded drill guide with depth gauge broke on an unknown procedure.The product threads were stripped and would not thread into the plate.It was unknown if there was surgical delay.Patient outcome and patient status was unknown.This complaint involves one (1) device.This report is 1 of 1 for (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, the 2.0mm threaded drill guide with depth gauge was broken.The product threads were stripped and would not thread into the plate.It is unknown if there were patient and surgical involvement.Upon manufacturer receipt of the device, it was noted that some of the marking paint was missing.Concomitant devices: trauma plate (part: unknown, lot: unknown, quantity: unknown).This report is for a 2.0mm threaded drill guide with depth gauge.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: a device history record (dhr) review was conducted: part: 323.061.Lot: 1382292.Manufacturing site: haegendorf.Release to warehouse date: 25.Aug.2005.The device history record shows this lot of (b)(4) pieces was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and functional criteria at the time of release with no issues documented during the manufacturing process.Review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.A product investigation was conducted.Visual inspection: the 2.0 mm threaded drill guide with depth gauge (part # 323.061, lot # 1382292, mfg # 25-aug-2005) was received at us cq with signs of wear.Parts of the orange coating was peeled off.There are signs of discoloration/rust.The drill guide showed no signs of being broken or having stripped threads.This is not consistent with the reported complaint condition.However, the complaint is confirmed for general signs of wear.Dimensional inspection: dimensional analysis was completed, the inner diameter of the distal threaded portion of the drill guide measured and is within specification based on relevant drawing.Document/specification review: the relevant drawing(s) was reviewed; review of the device history record showed that there were no issues during the manufacture of this product which would contribute to this complaint condition.Conclusion: the complaint condition is confirmed as the 2.0 mm threaded drill guide with depth gauge (part # 323.061, lot # 1382292) was received with signs of wear, discoloration and peeling of the orange coating.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces (such as being dropped, struck off axis or damaged during sterile processing).No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|