Investigation evaluation: an evaluation of the photos provided by the user of the device in the endoscopic view confirmed the report of incorrect cutting wire orientation.In the photos provided, the device exited the scope in the 11 o'clock direction.The distal end appeared to be twisted in the photos provided which could have lead to difficulty orientating the device.In the second photo provided, the cutting wire appears to have oriented in the 12 o'clock direction (appropriate orientation is approximately 11:00 - 1:00 o'clock).In both photos provided, we had difficulty determining the position of the papilla with respect to the distal end.An evaluation of the returned device could not confirm the report of incorrect cutting wire orientation but confirmed it based on the photos provided by the user.During our laboratory analysis, the sphincterotome was advanced through a duodenoscope that is placed in a simulated biliary position.The duodenoscope has an accessory channel that is 4.2 mm in diameter (model number olympus tjf-160v).The catheter exited the endoscope with the cutting wire facing 12 o¿clock.The device was then bowed and the cutting wire was facing 12 o'clock (appropriate orientation is approximately 11:00 - 1:00 o'clock).The sphincterotome catheter was subjected to a close visual examination and twisting of the tubing was observed at the distal end.A discrepancy or anomaly that could have contributed to the reported event was not observed during our laboratory analysis of the returned product.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Prior to distribution, all fusion omni-tome pre-loaded sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion omni-tome pre-loaded sphincterotome.The tip of sphincterotome was totally twisted, with wrong orientation [incorrect cutting wire orientation].The following information was also provided per the complaint form: the product had real bad orientation, not covering main feature of the device.They had to use a new one from our competitors.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|