Catalog Number 6021-0637
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Injury (2348); Joint Dislocation (2374); Reaction (2414)
Date of Event 09/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the product history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Patient was fixing a deck and experienced severe pain in his hip and could not move without pain.Update 10/september/2018: as cited in "adverse consequences details": patient required revision surgery.It was repeorted that the head disassociated from the stem.A considerable amount of black tissue was present in the joint.Pre-revision x-rays and an explant picture were provided.
 
Event or Problem Description
Patient was fixing a deck and experienced severe pain in his hip and could not move without pain.Update 10/september/2018: as cited in "adverse consequences details": patient required revision surgery.It was repeorted that the head disassociated from the stem.A considerable amount of black tissue was present in the joint.Pre-revision x-rays and an explant picture were provided.
 
Additional Manufacturer Narrative
Adding investigation conclusion; reported event: an event regarding disassociation & altr involving an accolade stem was reported.Disassociation was confirmed.Altr was not confirmed.Method & results: -product evaluation and results: the device was not returned however a photo was provided which noted wear on the trunnion of the stem and biological material on the coated segment.-medical records received and evaluation: a review of the provided medical records by a clinical physician was rejected stating [.] rejected for medical review by dr.Jaffe today - x-ray and photo confirm event of disassociation and trunnion erosion/corrosion; need operative reports, clinical/past medical history, examination of explanted components[.] -product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.-complaint history review: there were no other events for the lot provided.Conclusion: disassociation was confirmed by a clinician on review of the provided photos & x-ray however altr was not confirmed.Further information such as return of device, patient history, histopathology report & follow-up notes are needed to investigate this event further.If additional information and/or device becomes available, this investigation will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE PLUS TMZF HIP STEM #6
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7937984
Report Number0002249697-2018-03217
Device Sequence Number13937285
Product Code LPH
UDI-Device Identifier04546540510716
UDI-Public04546540510716
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2006
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2011
Device Catalogue Number6021-0637
Device Lot Number18035703
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/07/2018
Supplement Date Received by Manufacturer11/02/2018
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date11/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
-
-