DEPUY ORTHOPAEDICS, INC. 1818910 AML SM STATURE 12.0MM; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 155401120
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Swelling (2091); Discomfort (2330); Test Result (2695)
Date of Event 10/27/2015
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).Reporter is an attorney.
 
Event or Problem Description
Litigation alleges pain, discomfort, inflammation, swelling, difficulty ambulating, difficulty weight bearing, and added stem due to alleged elevated metal ions.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added : h6 product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: bt9ey1000.Device history batch: null.Device history review: 1 piece was placed on hold and split from the batch at operation 70 porocoat.This was per the approved local procedures at the time.No other deviations or anomalies were observed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AML SM STATURE 12.0MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7938026
Report Number1818910-2018-71424
Device Sequence Number770021
Product Code JDI
UDI-Device Identifier10603295058229
UDI-Public10603295058229
Combination Product (Y/N)N
Initial Reporter StateNY
PMA/510(K) Number
K012364
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2015
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/01/2017
Device Catalogue Number155401120
Device Lot NumberBT9EY1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer10/02/2019
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date10/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age57 YR
Patient Weight100
-
-