| Model Number |
MMT-551NAH |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that insulin pump had blank display.The blood glucose level was at 174 mg/dl the time of incident.Customer stated that the crack to the reservoir compartment.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with cracked reservoir tube window corner noted.Device passed displacement test, self test, unexpected restart error test, off no power test and all operating currents tested within the specific range.
|
| |
|