| Catalog Number |
L98113 |
| Medical Device Problem Codes |
Fracture (1260); Loose or Intermittent Connection (1371); Adverse Event Without Identified Device or Use Problem (2993); Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Scarring (2061); Inadequate Osseointegration (2646); No Code Available (3191)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Patient code: surgical intervention.
|
| |
|
Event or Problem Description
|
|
The corail revision prosthesis has been broken a couple of weeks/months after surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).
|
| |
|
Event or Problem Description
|
|
The corail revision shaft is broken after a few weeks, within 3 months after surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.After review of medical records, it was reported that the patient was revised to address fracture of the stem causing pain, leg length discrepancy and component loosening.Operative notes reported of neocapsule scarring and massive loosening was apparent which was pulled out with the bare hand.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: examination of the device confirmed the device had fractured.No anomalies in manufacturing records or complaint databases.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the dhr analysis of the batch indicated shows an initial conformance of this product with regards to its specification.For this batch, there was no deviation or non-conformance.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: examination of the device confirmed the device had fractured no anomalies in manufacturing records or complaint databases.Root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: he dhr analysis of the batch indicated shows an initial conformance of this product with regards to its specification.For this batch, there was no deviation or non-conformance.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|