COOK INC TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER; DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number HNB5.0-35-100-P-8S-NEFF
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) number: pre-amendment.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the tip of a torcon nb advantage angiographic catheter was found to be kinked and broken upon removal from the package.This device was not used.Another torcon nb advantage angiographic catheter was opened and used in the unspecified procedure.No patient adverse effect has been reported.
 
Additional Manufacturer Narrative
A review of the complaint history, device history record, drawing, instructions for use (ifu), manufacturing instructions, quality control, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a bend on the device material, which was located at the 5th sideport from tip of the device.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be established.We will continue to monitor for similar complaints.Per the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORCON NB ADVANTAGE ANGIOGRAPHIC CATHETER
Common Device Name
DQO CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7938356
Report Number1820334-2018-02609
Device Sequence Number2305094
Product Code DQO
UDI-Device Identifier00827002110267
UDI-Public(01)00827002110267(17)210606(10)NS8880529
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/06/2021
Device Catalogue NumberHNB5.0-35-100-P-8S-NEFF
Device Lot NumberNS8880529
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/20/2018
Supplement Date Received by Manufacturer01/02/2019
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date01/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-