• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNITRAX C-TAPER SLEEVE +0MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNITRAX C-TAPER SLEEVE +0MM; PROSTHESIS, HIP, HEMI-, FEMORAL, METAL Back to Search Results
Catalog Number 6942-7-065
Medical Device Problem Codes Fitting Problem (2183); Positioning Problem (3009)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/07/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
When completed, the evaluation summary will be submitted in a supplemental report.
 
Event or Problem Description
Primary procedure, hip.It was reported the unitrax c-taper sleeve would not stay on the stem when impacted on.Head remained impacted on the sleeve.The construct was removed and a c-taper lfit head was implanted instead.There was a surgical delay of approximately 10 minutes.
 
Additional Manufacturer Narrative
An event regarding a size/fit issue involving a hip sleeve was reported.The event was not confirmed.Dimensional & functional inspection: not performed as there is evidence from the visual inspection that the device is damaged.Material analysis: material analysis was not performed due to the nature of the event medical records received and evaluation: no medical records were received for review with a clinical consultant device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports are needed to fully investigate the event.If further information becomes available, this investigation will be re-opened.
 
Event or Problem Description
Primary procedure, hip.It was reported the unitrax c-taper sleeve would not stay on the stem when impacted on.Head remained impacted on the sleeve.The construct was removed and a c-taper lfit head was implanted instead.There was a surgical delay of approximately 10 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNITRAX C-TAPER SLEEVE +0MM
Common Device Name
PROSTHESIS, HIP, HEMI-, FEMORAL, METAL
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7938396
Report Number0002249697-2018-03222
Device Sequence Number1346287
Product Code KWL
UDI-Device Identifier07613327013467
UDI-Public07613327013467
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K992570
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Catalogue Number6942-7-065
Device Lot Number64998703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2018
Initial Date Received by Manufacturer 09/07/2018
Supplement Date Received by Manufacturer10/30/2018
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date11/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
-
-