Model Number N/A |
Device Problems
Contamination /Decontamination Problem (2895); Packaging Problem (3007)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Expected, not yet received.
|
|
Event Description
|
It was reported that during incoming inspection at the warehouse, it was discovered that ten products were found to have the seal peeling off and a stain found on the product.Further, there were eight products that were found to have stains on the products only.There was no patient involvement.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.For passivation, the guide wires are placed on a rack with two or more contact points.These contact points cause a small portion of the guide wire to not receive a full passivation coating, leading to an inconsistent coating appearing as a "stain" or mark.These stains are not corrosion, and do not impact the fit, form, or function of the guidewires; they are merely cosmetic.Device analysis indicated that the device met specification.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information received.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Corrected: remove device code "packaging problem" as it is no longer applicable for this complaint.
|
|
Event Description
|
It was reported that during incoming inspection at the warehouse, it was discovered that eighteen products were found to have a stain found on the product.There was no patient involvement.
|
|
Search Alerts/Recalls
|