• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; MAIN BODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX (TRIVASCULAR INC.) OVATION IX; MAIN BODY Back to Search Results
Model Number TV-AB2380-J
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 07/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.At the completion of the clinical assessment, clinical was able to find substantial evidence to support the following event of a type 1a endoleak of the main body.Procedure related harms, device, user, procedure or anatomy relatedness of this complaint could not be determined.The final patient status was reported to be in stable post the secondary endovascular procedure endologix continues to investigate this event and similar events to ensure the highest quality and patient safety.
 
Event or Problem Description
At the completion of the clinical assessment on 09/05/2018 for a pervious report event (3008011247-2018-00168), clinical was able to find substantial evidence to support the following additional findings of a type 1a endoleak of the main body.This report is for the type 1a endoleak of the main body only.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVATION IX
Common Device Name
MAIN BODY
Manufacturer (Section D)
ENDOLOGIX (TRIVASCULAR INC.)
3910 brickway blvd.
santa rosa CA 95403
MDR Report Key7939064
Report Number3008011247-2018-00214
Device Sequence Number1797743
Product Code MIH
UDI-Device IdentifierM701TVIL1422120J1
UDI-Public+M701TVIL1422120J1/$$3210330FS032118418
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
P120006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/12/2021
Device Model NumberTV-AB2380-J
Device Catalogue NumberTV-IL1422120-J
Device Lot NumberFS022218-45
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/05/2018
Initial Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ILIAC LIMB, LOT # FS032118-41; ILIAC LIMB, LOT # FS032918-66
Outcome Attributed to Adverse Event Required Intervention;
-
-