| Model Number |
1600US |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Product event summary: the battery charger was returned for evaluation.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned battery charger revealed that the unit passed visual inspection and functional testing.Supplemental testing revealed that the battery charger was assembled with incorrect inductors.A thermal reading was performed between (b)(4) and the same battery charger with corrected inductors.The results revealed a slight increase in temperature.The incorrect inductors causes an increase in power consumption across transistors in the charging circuit, increasing the temperature surrounding the printed circuit board (pcb).The increased temperature is likely related to the reported overheating of the battery charger.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to improper assembly.An internal investigation was initiated to investigate battery chargers assembled with incorrect inductors.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported, by product analysis,that the incorrect inductor was used during the assembly process.The battery charger was exchange.No patient complications have been reported as a result of this event.
|
| |
|
Search Alerts/Recalls
|