• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MENISCAL DEPLOYMENT GUN; ORTHOPAEDIC CERCLAGE APPLIER Back to Search Results
Catalog Number 228143
Device Problems Positioning Failure (1158); Mechanical Jam (2983)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported by the customer in (b)(6) that the gun did not fire and needle got stuck.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).
 
Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device was received and evaluated.Initial visual observation of the device revealed the deployment slider appeared to be bent at the distal tip.The gray trigger could be triggered properly and the deployment slider protrudes out fully bent.The red trigger rod was present and functioned normally.This bending of the key feature is typical of an extreme load applied to the needle as would occur in a misuse condition.The bending could also result from the use error of improper assembly of the needle onto the applier.The bending could be the cause for the needle to get stuck and gun not firing as intended.A non-conformance search was performed for this part (228143), lot (l213225) combination and no non-conformance's were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Type of Device
ORTHOPAEDIC CERCLAGE APPLIER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7939208
MDR Text Key123373037
Report Number1221934-2018-54832
Device Sequence Number1
Product Code GEF
UDI-Device Identifier10886705010059
UDI-Public10886705010059
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Catalogue Number228143
Device Lot NumberL213225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/22/2018
Date Manufacturer Received11/12/2018
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-