It was reported a company representative, that a generator immediately did not look right to the surgeon after being taken out of the packaging for an implant surgery.The septum plug looked frayed and did not look right initially the surgeon could not get the lead to go past the first connector block, but eventually he was able to use the hex screw driver to get the lead in.Then the lead pin tip would not go past the second connector block.The surgeon indicated that the generator and lead connection didn't look right.Impedance was ok at 2030 ohms but they had hesitations about using the generator.The surgeon had reportedly done many surgeries in the past.It was noted that he had tried irrigating the lead cavity.The attending company representative indicated that the surgeon probably tried to use the hexscrew 20 times.The surgeon decided to use a new generator.The manufacturer's device history records of the suspect generator were reviewed.The generator passed final quality and functional specification prior to release.The suspect product was received but product analysis on it has not been completed to date.No further relevant information has been received to date.
|
Product analysis was completed on the returned generator.The septum was cored and the setscrew showed mechanical wear at the socket, suggesting numerous insertions with a torque wrench.In addition, the septum showed damage on the underneath side, which indicates that the setscrew was extracted up into the septum.The reported allegation of ¿insertion difficulties¿, was duplicated in the pa lab, due to the debris (septum plug material and possible adhesive) in the header lead cavity.The in-line cavity go gauge test, designed to verify proper lead cavity dimensions in the header area, passed.However, a bench lead inserted into the pulse generator header, would not go past the negative connector block (lab conditions), due to septum plug debris and possible adhesive debris in the header lead cavity.Eventually, the septum and possible adhesive debris was compacted enough that the bench lead pin barely past the negative connector block (lab conditions).Note that the septum plug debris may have resulted from multiple insertion attempts where the septum plug was cored it is unclear how the possible adhesive was found to be in the cavity in that is unknown if that material was left over from the manufacturing process or it was loosened and then dislodged during the insertion attempts.Other than the visual anomalies observed, there were no additional performance or any other type of adverse conditions found with the pulse generator.The generator passed functional specifications per the final electrical test.Note that product analysis is unable to assess the conditions of generators out of the packaging after they have been used.
|