KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG LEVEL ONE CMF; PLATE Back to Search Results
Model Number 25-314-85-91
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Scar Tissue (2060)
Date of Event 09/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: mdr: 9610905-2018-00248.Reference exemption number e2017029.
 
Event or Problem Description
It was reported the plates were sitting proud.The plates were removed.
 
Additional Manufacturer Narrative
An investigation was performed using a stereo microscope revealed surface damage and bending but no breakage.Further observation determined there were no indications of material or manufacturing defects.The risk and device history records were not reviewed as there was no failure found.The investigation results conclude that the root cause is patient related as no device issue was found.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
PLATE
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strasse 10
muehlheim/donau, 78570
GM  78570
MDR Report Key7939736
Report Number9610905-2018-00247
Device Sequence Number11339173
Product Code JEY
UDI-Device Identifier00888118038722
UDI-Public(01)00888118038722
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K944565
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2018
Report Date (Section F) 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number25-314-85-91
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Returned to Manufacturer09/28/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date10/25/2018
Event Location Hospital
Date Report to Manufacturer09/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer10/25/2018
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date11/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age21 YR
Patient Weight60
-
-