| Model Number |
MT22495-BLU |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the receiver ceased to function.No additional event or patient information is available.The device has been received for investigation.A follow-up will be submitted upon completion of device investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Com-(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An external visual inspection was performed and passed.A receiver charge and boot was performed and passed.A receiver download was performed and an unexpected receiver shutdown was observed.A functional test was performed and passed relevant testing.The receiver case was opened, and an internal visual inspection was performed and failed due to a damaged battery.Confirmation of the complaint was confirmed.The probable cause was a damaged receiver battery.
|
| |
|