MEDTRONIC, INC. DEFIB EPI PATCH; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number 6721L-50
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Positioning Problem (3009)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 6721m-50 lead, implanted: (b)(6) 2016; 4968-60 lead, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) battery depleted prematurely.The device was explanted and replaced.It was also reported that the two epicardial defibrillation patches dislodged as they appeared to have migrated away from the heart towards the diaphragm.The patches were capped.The surgeon was unable to retrieve the dislodged patches as one had migrated to a trough around the diaphragm and one had scarred over below and behind the heart.A subcutaneous coil was implanted and successfully converted the patient during defibrillation testing.The coil remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFIB EPI PATCH
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key7940104
Report Number2182208-2018-01808
Device Sequence Number19262954
Product Code LWS
UDI-Device Identifier00681490148733
UDI-Public00681490148733
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P900061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/11/2018
Device Model Number6721L-50
Device Catalogue Number6721L-50
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2018
Initial Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
511212 LEAD (X2), 511211 LEAD (X2), 6721L-50 LEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient Weight125
-
-