Model Number C12787
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
The black power cord connected to the quantien monitor began to smoke.The procedure was immediately stopped, and the monitor was turned off.The procedure was finished with a non-abbott device.Patient's information (e.G.Age, weight, gender, ethnicity, race, patient's initials) cannot be handled by abbott in absence of patient's written consent as required by the personal data protection national legislation.National legislation prevents the recording of such information.A written consent has not been obtained in this case; therefore, this information is not available.
 
Additional Manufacturer Narrative
Based on the information provided to abbott and the investigation performed, the root cause of the reported event cannot be conclusively determined as no abnormalities were identified.The returned product operated as designed during the evaluation period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTIEN
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key7940394
Report Number2184149-2018-00167
Device Sequence Number2014595
Product Code DQK
UDI-Device Identifier05415067000996
UDI-Public05415067000996
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K123984
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC12787
Device Catalogue NumberC12787
Device Lot Number6049866
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/25/2018
Supplement Date Received by Manufacturer01/04/2019
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date01/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-