• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125275-23
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Angina (1710); Death (1802); Cardiac Enzyme Elevation (1838); Stenosis (2263)
Date of Event 08/13/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effects of angina and stenosis are listed in the xience alpine, everolimus eluting coronary stent system (eecss), instructions for use as known patient effects of coronary stenting procedures.A conclusive cause for the reported patient effects and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
It was reported that on 03/07/2018, the patient underwent a coronary stenting procedure with implantation of a 2.75 x 23 mm xience alpine stent in the mid right coronary artery (rca).On (b)(6) 2018, the patient began to experience chest discomfort.The patient was re-hospitalized on (b)(6) 2018.Cardiac enzymes were elevated; however, electrocardiogram noted no changes suggestive of myocardial infarction.Coronary catheterization was performed and noted in-stent restenosis.Coronary intervention was performed and the event resolved on (b)(6) 2018.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).The reported patient effect of death is listed in the xience alpine, everolimus eluting coronary stent system (eecss), instructions for use (ifu) as known patient effects of coronary stenting procedures.A conclusive cause for the reported death and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Additional Manufacturer Narrative
(b)(4).Correction: type of reportable event.
 
Event or Problem Description
Subsequent to the initial 30-day medwatch report, the following information was received: on (b)(6) 2018, the patient died at home.No intervention was performed.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7940673
Report Number2024168-2018-07724
Device Sequence Number11247013
Product Code NIQ
UDI-Device Identifier08717648199271
UDI-Public08717648199271
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source health professional,study
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/05/2020
Device Catalogue Number1125275-23
Device Lot Number7092641
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer02/14/2019
03/19/2019
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date03/11/2019
03/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization; Required Intervention;
Patient Age74 YR
Patient Weight63
-
-