MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION TIGHTRAIL ROTATING DILATOR SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION TIGHTRAIL ROTATING DILATOR SHEATH Back to Search Results
Model Number 545-513
Medical Device Problem Codes Failure to Align (2522); Material Deformation (2976); Mechanical Jam (2983); Noise, Audible (3273); Physical Resistance/Sticking (4012); Sharp Edges (4013)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient information was requested from the site, but not provided.During evaluation of the device it was found that the first 3-4 inches of the tightrail sheath was flexible.Biologic material was visible on the outside of the device with a large amount observed at blades.The tri-coil was herniated within the device.The handle was in the full forward position with the blades extended.The handle was disassembled.Inside, the covers and trigger tabs were in excellent condition, showing little use.No biologic material was present inside of handle.The trigger pin was broken out of the trigger causing a rattling noise.Constant force spring showed little wear.Rear strain relief was in good condition.There was a groove cut into the barrel, with multiple areas of track damage from the trigger pin.This is evidence that the trigger pin jumped out of barrel channel and damaged the sleeve.The sleeve was deformed from the trigger pin.The far track corner was sheared off and lateral grooves were present on the outside of the sleeve.The sleeve tabs were in proper position, but the home track corner was broken.The drive unit would not actuate by hand as the tri-coil was herniated causing an abnormal noise.
 
Event or Problem Description
A philips representative reported that during a cardiac lead management procedure on (b)(6) 2018, the handle of the tightrail 545-513 device malfunctioned.The blades were not rotating, and the handle made a clicking noise.No patient harm occurred.It was reported that this was a right sided entry into the body, with extreme calcium lesions.The suspect tightrail 545-513 device was returned for evaluation and it was found during evaluation on 14sep2018, that the inner coil was herniated.The herniation caused the blades to be stuck in the extended position.These evaluation findings made the manufacturer aware that this issue could lead to patient injury if it were to recur, as the blades extending through the tip of the device could lead to unintended tissue damage and/or emboli.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIGHTRAIL ROTATING DILATOR SHEATH
Common Device Name
TIGHTRAIL
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key7940901
Report Number1721279-2018-00137
Device Sequence Number16376488
Product Code DRE
UDI-Device Identifier00813132020897
UDI-Public00813132020897
Combination Product (Y/N)Y
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K150360
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/10/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2019
Device Model Number545-513
Device Catalogue Number545-513
Device Lot NumberFRJ17L30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/14/2018
Initial Report FDA Received Date10/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/01/2005
-
-