• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL MINI ROTATING DILATOR SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL MINI ROTATING DILATOR SHEATH Back to Search Results
Model Number 540-011
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Perforation of Vessels (2135); Iatrogenic Source (2498)
Date of Event 09/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient's weight was unavailable from the site.Lot # and expiration date could not be determined as the site discarded the device.Manufacture date could not be determined since the lot# is unavailable.Site discarded the device.Therefore, device evaluation cannot be completed.There is no allegation that the device malfunctioned.
 
Event or Problem Description
A philips representative reported that during a cardiac lead management procedure 5 leads were to be extracted.The physician utilized the 12fr glidelight laser sheath first, the 9fr long tightrail rotating dilator sheath second, and then an 11fr tightrail mini rotating dilator sheath 540-011 with a lead locking device (lld).After the right atrium tendril lead was extracted utilizing the spectranetics tightrail mini 540-011, the patient's blood pressure dropped significantly.An effusion was visible on imaging.There was an injury to the superior vena cava (svc)/right atrial (ra) junction.Rescue efforts commenced and were successful.Patient survived the procedure and all 5 leads intended for removal were successfully extracted.
 
Additional Manufacturer Narrative
It was detected on (b)(6) 2018 that on the initial medical device report, was populated incorrectly as (b)(6) 2018, but should have been (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TIGHTRAIL MINI ROTATING DILATOR SHEATH
Common Device Name
TIGHTRAIL
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key7940904
Report Number1721279-2018-00144
Device Sequence Number13789686
Product Code DRE
UDI-Device Identifier00813132020866
UDI-Public00813132020866
Combination Product (Y/N)Y
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K150360
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/12/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number540-011
Device Catalogue Number540-011
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer10/08/2018
Initial Report FDA Received Date10/05/2018
Supplement Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SPECTRANETICS CVX-300 LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH; SPECTRANETICS LEAD LOCKING DEVICE (LLD); SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age70 YR
Date Report Sent to FDA01/01/2005
-
-