| Model Number |
1000 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio-control also observed that the device was delivering monophasic energy rather than biphasic energy.The energy delivered by the device was also at a lower energy level than what was selected.The cause of the observed issue was determined to be due to a therapy harness ferrite bead being set over an integrated circuit, designator u1, in error during the manufacturing of the device.The incorrectly set therapy harness ferrite bead broke u1 legs 1 and 16 coming from the analog pcb assembly.U1 leg 16 no longer had a connection, which resulted in the device delivering a monophasic shock at a reduced energy level.The customer received a replacement device.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report a non-critical issue with their device.There was no patient use associated with the reported event.Upon evaluation of the customer's device, physio-control observed that the device was delivering monophasic energy rather than biphasic energy.The energy delivered by the device was also at a lower energy level than what was selected.As a result, the incorrect defibrillation therapy may be delivered, if it was needed.
|
| |
|