| Model Number |
DADE INNOVIN |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer contacted a siemens customer care center (ccc) and reported that a discordant, falsely elevated prothrombin time (pt) and pt international normalized ratio (inr) results were obtained on patient samples on a sysmex ca-620 system using dade innovin reagent.The customer also reported that there were no error flags generated by the system and quality controls (qcs) were within the acceptable range.Siemens discussed how to properly handle samples with the customer.A siemens customer service engineer (cse) was dispatched to the customer's site and performed a scheduled preventative maintenance.The customer ran a precision test using quality controls (qcs), resulting acceptably.Customer also ran four samples for troubleshooting purposes with no issues noted.No instrument malfunction was identified.Siemens further investigated the issue and determined that the cause of the discordant results was potentially related to sample handling practices, poor sample integrity or pre-analytical variables.No further evaluation of this device is required.The systems and reagents are performing according to specifications.Mdr 9610806-2018-00102 was filed for the same issue.
|
| |
|
Event or Problem Description
|
|
Discordant, falsely elevated prothrombin time (pt) and pt international normalized ratio (inr) results were obtained on two patient samples on a sysmex ca-620 system, using the dade innovin reagent.The discordant results were not reported to the physician(s).Sample id (b)(6) was respun and then both samples were repeated on the same system, using the same reagent, multiple times.The repeat result of 29.5 pt and 2.98 pt inr for sample id (b)(6) and the result of 41.7 pt and 4.25 pt inr for sample id (b)(6) were reported, as the correct results to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant pt and pt inr patient results.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information (12-october-2018): the customer clarified that the patient of sample id (b)(6) was on anticoagulation medication (coumadin).Corrected information (12-oct-2018): upon further investigation, siemens determined that the initial results for sample id (b)(6) was 50.2 seconds and 5.07 inr, and not 90.2 seconds and 9.20 inr.
|
| |
|
Search Alerts/Recalls
|