Model Number MICL 13.2 -13.5
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient code: (b)(4) - secondary surgical intervention, explant.Word order search: no similar complaint type events were reported for units within the same lot.(b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon inserted a 13.2 mm micl13.2, -13.5 diopter, implantable collamer lens in patients right eye (od) on (b)(6) 2018.The lens remained folded and when it unfolded, the haptic were in the ac angle (lens flipped), the surgeon was unable to view the orientation dot of lens.On (b)(6) 2018 the lens was removed and a same type model and length lens was implanted.Patient is doing very well.
 
Additional Manufacturer Narrative
Device evaluation: lens was returned in vial, in liquid.Visual inspection found no visible damage to lens.(b)(4).
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key7943035
Report Number2023826-2018-01476
Device Sequence Number941026
Product Code MTA
UDI-Device Identifier00841542103367
UDI-Public00841542103367
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2019
Device Model NumberMICL 13.2 -13.5
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Initial Date Received by Manufacturer 09/25/2018
Supplement Date Received by Manufacturer10/17/2018
11/13/2018
Initial Report FDA Received Date10/08/2018
Supplement Report FDA Received Date10/18/2018
11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
-
-