MAUDE Adverse Event Report: ARTHREX, INC. UNI GLENOID-PERIPHERAL LOCK SCRW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNI GLENOID-PERIPHERAL LOCK SCRW; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number UNI GLENOID-PERIPHERAL LOCK SCRW
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a rtsa procedure, the peripheral locking screw, ar-9145-36, broke during insertion into the glenoid baseplate.What was left of the screw in the glenoid was left in order to not fracture the glenoid.The broken piece was secured in the baseplate.
 
Additional Manufacturer Narrative
Complaint confirmed.Visual evaluation of the returned ar-9145-36 revealed that the device had sheared off at the second, proximal thread prior to receipt for investigation.The most likely cause of this event is not implanting the screw coaxial to the hole, or bending/prying/leveraging the screw during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI GLENOID-PERIPHERAL LOCK SCRW
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7943045
Report Number1220246-2018-00691
Device Sequence Number857061
Product Code PHX
UDI-Device Identifier00888867057739
UDI-Public00888867057739
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2023
Device Model NumberUNI GLENOID-PERIPHERAL LOCK SCRW
Device Catalogue NumberAR-9145-36
Device Lot Number18.00132
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 09/17/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date10/08/2018
Supplement Report FDA Received Date10/26/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-