Model Number 638RL28
Medical Device Problem Code Incomplete Coaptation (2507)
Health Effect - Clinical Codes Mitral Regurgitation (1964); Mitral Valve Stenosis (1965)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that eight years and eleven months post implant of this 28mm mitral annuloplasty ring, the ring was explanted and replaced with a 27mm bioprosthetic valve.The indication for replacement was severe regurgitation and moderate stenosis.Examination of the native mitral valve revealed thickened anterior and posterior leaflets.The posterior leaflet appeared retracted, and the surgeon felt this was contributing to the insufficiency.During the replacement procedure, a trapezoid portion of the anterior leaflet was resected, and the posterior leaflet was retained.Additionally, the native tricuspid valve was repaired with an annuloplasty ring.Post-operative transesophageal echocardiogram (tee) demonstrated mild mitral insufficiency and trivial tricuspid insufficiency.No additional interventions took place, and no additional adverse patient effects were reported. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CG FUTURE BAND
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVE DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7943244
Report Number2025587-2018-02674
Device Sequence Number19168344
Product Code KRH
UDI-Device Identifier00613994521057
UDI-Public00613994521057
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K061127
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/24/2014
Device Model Number638RL28
Device Catalogue Number638RL28
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2018
Initial Report FDA Received Date10/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient Weight77
-
-