• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/8 HOLES/LEFT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/8 HOLES/LEFT; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.118.207
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional patient identifier: (b)(6).Additional product code: hwc.Date of implant reported as (b)(6) 2018, exact date is not known.Dhr review was completed.Part# 02.118.207, lot# h477525, mfg date: 24 october 2017, mfg location: (b)(6), expiration date: n/a, noted nonconformances: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of 2.7/3.5mm va-lcp anterolateral distal tibia product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A dhr review for raw material p/n 14089 is not required as the product was not broken this lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported patient underwent an ankle procedure in (b)(6) 2018 to implant two (2) locking compression plate (lcp) one third tubular plates, one (1) 2.7/3.5mm variable angle locking compression plate (va-lcp) anterolateral distal tibia plate, one (1) 2.7/3.5mm variable angle locking compression plate (va-lcp) medial distal tibia plate, thirteen (13) 3.5mm cortex screws, twelve (12) 2.7mm va locking screws, one (1) 7.0mm washer, one (1) 4.0mm cannulated screw, and one (1) 4.0mm cancellous screw.Patient was returned to surgery on september 17, 2018 where all hardware was removed due to infection, inflammation, and delayed healing.Surgery was completed successfully with no delay.Patient status reported as okay.Patient is reported to be incarcerated and non-compliant.This report is for one (1) 2.7/3.5mm va lcp anterolateral distal tibia plate.This is report 3 of 9 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/8 HOLES/LEFT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7944321
Report Number2939274-2018-54226
Device Sequence Number19397016
Product Code HRS
UDI-Device Identifier10886982038401
UDI-Public(01)10886982038401
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K121601
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.118.207
Device Catalogue Number02.118.207
Device Lot NumberH477525
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/17/2018
Initial Report FDA Received Date10/08/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient Weight72
-
-