• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #5 L-CEM; KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #5 L-CEM; KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS Back to Search Results
Catalog Number 5510F501
Medical Device Problem Codes Osseointegration Problem (3003); No Apparent Adverse Event (3189)
Health Effect - Clinical Codes Injury (2348); Ambulation or Postural Difficulties (2544); Inadequate Osseointegration (2646); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/26/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event or Problem Description
Patient called stating she had a left knee replacement about a year ago.Few months after implantation she started experiencing loosening and difficulty walking.Patient was revised.
 
Event or Problem Description
Patient called stating she had a left knee replacement about a year ago.Few months after implantation she started experiencing loosening and difficulty walking.Patient was revised to competitive devices on (b)(6) 2018.Patient is seeking compensation.Update per medical review: patient had a right knee revision on (b)(6) 2018 (competitor devices), not a left knee revision.This pi to cover "loosening & difficulty walking" without revision.
 
Additional Manufacturer Narrative
Additional information: date of birth correction: removed explant date as device remains implanted.An event regarding alleged loosening & difficulty walking involving a triathlon femoral component was reported.The event was not confirmed.Method & results: -product evaluation and results: not performed as the device was not returned as it remains implanted.-medical records received and evaluation: review of medical records by a consulting clinician revealed: "on (b)(6) 2017 a primary left total knee arthroplasty was performed for a pre- and post- operative diagnosis of ¿left knee arthritis and deformity¿.No primary operative report of the right total knee stated to be performed in (b)(6) 2015 or listing of the primary components removed at revision surgery to competitive products on (b)(6) 2018 is available.There is no documentation of any clinical problems with the left stryker total knee arthroplasty implanted on (b)(6), 2017, which was not revised.There are no x-rays available for review or examination of explanted right total knee arthroplasty components to determine if they were stryker.In this morbidly obese chronic pain patient dependent on narcotics, no documented complaints referable to the stryker primary total knee arthroplasty performed on (b)(6) 2017 in the left knee are noted.The event description appears to confuse the left total knee arthroplasty with the right revision surgery." -product history review: review indicated all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: it was reported that allegedly the patient experienced loosening and difficulty walking.The medical review noted: "there is no documentation of any clinical problems with the left stryker total knee arthroplasty implanted on (b)(6) 2017, which was not revised." the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation as they remain implanted and insufficient medical information was provided.If the devices and/or additional information are received, this investigation will be reopened and re-evaluated.
 
Additional Manufacturer Narrative
Correction - january 25, 2024: as per the medical review, "there is no documentation of any clinical problems with the left stryker total knee arthroplasty implanted on (b)(6) 2017, which was not revised.There are no x-rays available for review or examination of explanted right total knee arthroplasty components to determine if they were stryker." additionally, the patient corrected her statement to report problems with her right knee, not the left stryker knee.Therefore, there is no indication that this patient's experience is related to stryker devices and this complaint is being canceled.
 
Event or Problem Description
Update from patient: patient stated she had problems with her right knee not left knee.She stated she will send her medical records.Update from medical review: there is no documentation of any clinical problems with the left stryker total knee arthroplasty implanted on (b)(6) 2017, which was not revised.There are no x-rays available for review or examination of explanted right total knee arthroplasty components to determine if they were stryker.Patient called stating she had a left knee replacement about a year ago.Few months after implantation she started experiencing loosening and difficulty walking.Patient was revised to competitive devices on (b)(6) 2018.Patient is seeking compensation.Update per medical review: patient had a right knee revision on (b)(6) 2018 (competitor devices), not a left knee revision.This pi to cover allegation of "loosening & difficulty walking" of an unknown left knee without revision.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #5 L-CEM
Common Device Name
KNEE JOINT PATELLOFEMOROTIBIAL METAL/POLYMER POROUS-COATED UNCEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key7944484
Report Number0002249697-2018-03229
Device Sequence Number1343640
Product Code MBH
UDI-Device Identifier07613327040296
UDI-Public07613327040296
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Catalogue Number5510F501
Device Lot NumberA7L4R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer07/29/2019
01/25/2024
Initial Report FDA Received Date10/08/2018
Supplement Report FDA Received Date08/23/2019
02/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/14/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexFemale
Patient Weight118 KG
-
-