An event regarding disassociation involving a metal head and an accolade stem was reported.The event was confirmed through clinician review of the medical records provided.Method & results: product evaluation and results: visual inspection was performed as part of the material analysis report (mar), which indicated: damages were observed on the neck of the stem, consistent with wear mechanisms due to the cyclic contact between the head taper and stem trunnion after the loss of their taper lock.Damage consistent with the explantation process in addition to biological fixation were also observed on the stem.Debris was also observed on the stem.A material analysis has been performed.The report concluded: damage was observed on the stem trunnion and head taper.This damage was consistent with wear mechanisms due to the cyclic contact between the head taper and stem trunnion after the loss of their taper lock.Debris was also observed on the stem trunnion.Eds showed the stem was consistent with astm f1813 alloy, the head was consistent with astm f1537 alloy and the debris was consistent with a corrosion product, stem base alloy and biological material.Damage consistent with impingement was observed on the distal rim of the insert.Scratching was observed on the insert, which is a common damage mode of uhmwpe.Yellow discoloration consistent with absorption of synovial fluid was also observed on the insert.Clinician review: a review of the provided medical records by a clinical consultant indicated: cup malposition in low inclination with additional inappropriate use of a 10° cup liner in a cup shell with already low inclination has contributed to an overload condition in the arthroplasty leading to excessive corrosion in the taper junction with gradual loss of taper lock functionality and ultimately disassociation between femoral head and trunnion requiring revision surgery.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: an event regarding disassociation between an accolade stem and metal head was reported.Visual inspection was performed as part of the material analysis report (mar), which indicated: damages were observed on the neck of the stem, consistent with wear mechanisms due to the cyclic contact between the head taper and stem trunnion after the loss of their taper lock.Damage consistent with the explantation process in addition to biological fixation were also observed on the stem.Debris was also observed on the stem.A material analysis has been performed.The report concluded: damage was observed on the stem trunnion and head taper.This damage was consistent with wear mechanisms due to the cyclic contact between the head taper and stem trunnion after the loss of their taper lock.Debris was also observed on the stem trunnion.Eds showed the stem was consistent with astm f1813 alloy, the head was consistent with astm f1537 alloy and the debris was consistent with a corrosion product, stem base alloy and biological material.Damage consistent with impingement was observed on the distal rim of the insert.Scratching was observed on the insert, which is a common damage mode of uhmwpe.Yellow discoloration consistent with absorption of synovial fluid was also observed on the insert.The medical review revealed: cup malposition in low inclination with additional inappropriate use of a 10° cup liner in a cup shell with already low inclination has contributed to an overload condition in the arthroplasty leading to excessive corrosion in the taper junction with gradual loss of taper lock functionality and ultimately disassociation between femoral head and trunnion requiring revision surgery.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|