| Model Number |
1506-10-010 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Edema (1820); Bone Fracture(s) (1870); Pain (1994); Swelling (2091); Weakness (2145); No Code Available (3191)
|
| Date of Event |
06/19/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Occupation: attorney.
|
| |
|
Event or Problem Description
|
|
Right revision to address pain, instability, quad atrophy and quad weakness, effusion, and loosening of the tibial component at the implant/cement interface.There was wear with grooving and pitting to the underside of the rotating platform.There were cracks in the cement and a small crack in the medial condylar region, not complete, but consistent with probable stress fracture.Patella was noted to be overstuffed.
|
| |
|
Additional Manufacturer Narrative
|
|
Udi: (b)(4).(b)(4) was used to capture medical device removal and surgical intervention.
|
| |
|
Event or Problem Description
|
|
(b)(4) and medical records alleges tibial loosening at the bone to cement interface.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Added: h6 (patient).H6 patient code: no code available (3191) was used to capture joint instability and atrophy.
|
| |
|