The customer reported false a positive fy(a) antigen typing result when testing with seraclone anti-fy(a).The customer stated that the donor sample was typed negative with immucor hea dna testing.The customer did not return the supposedly defective product for investigational testing, but he returned the donor sample that had caused a false positive test result.Therefore our quality control laboratory tested their seraclone anti-fya retention sample with the donor sample and could confirm the customer's finding.Subsequently the donor sample was tested in the direct antiglobulin test (dat) and received a positive result due to a sensitization with igg.The igg antibody was removed from the donor red blood cells by a heat elution method, following a retest of the donor sample with the supposedly defective product.Now the donor sample reacted correctly negative.Furthermore the supposedly defective product was tested for potency and specificity.All acceptance criteria were met.The minimum titer of 16 was exceeded.All positive and negative reactions were correct.We did not observe any false positive reaction.The cause of the false positive result of the specific donor sample with the supposedly defective product was the igg sensitization, confirmed by the positive dat.In the section limitation of the instruction for use there is an appropriate note: "samples with a positive direct antiglobulin test, cold agglutinins, or rouleaux formation may show false positive results in testing with monoclonal antibodies." testing by our quality control laboratory confirmed that the allegedly defective lot of seraclone anti-fy(a) functions correctly.A review of the batch record documentation showed no irregularities which might have negatively affected the quality of the allegedly defective lot.
|