| Catalog Number |
UH1-48-26 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Injury (2348); Joint Dislocation (2374)
|
| Date of Event |
09/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event regarding periprosthetic fracture involving a uhr head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned medical records received and evaluation: no medical records were received for review with a clinical consultant device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.Not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's left hip was revised approximately 1 month post-op due to recurrent dislocations.A stryker bipolar component and femoral head as part of a hemi hip construct were revised to a total hip with competitor acetabular components and another stryker head.The other femoral components were not revised.Rep reported that x-rays, medical records, and additional information are not available due to hospital policy.Update: surgeon reported cause of dislocation was due to a posterior acetabular rim fracture (no trauma reported to rep).Surgeon also reported that "component placement was excellent".
|
| |
|
Search Alerts/Recalls
|