MAUDE Adverse Event Report: DEPUY MITEK LLC US SCR/WASH HEXDRV 2.5MM *EA; SURGICAL SCREWDRIVER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US SCR/WASH HEXDRV 2.5MM *EA; SURGICAL SCREWDRIVER, REUSABLE Back to Search Results
Catalog Number 219124
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4)-incomplete.The lot number is currently unavailable.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the sales rep via phone that during a knee surgery procedure, it was observed that the screw washer hex driver tip broke while putting in the screw.The sales rep stated that no fragments were created.The sales rep stated that there were no procedural or anatomical factors that contributed to breakage.The case was completed with another like-device with no patient harm or delays.The sales rep was not present for the case and could not provide any additional information.The sales rep was unable to provide a lot number.The device is being returned for evaluation.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by dfa.Depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch.A follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Visual observation confirms the distal tip of the driver is broken but not returned with the device.Hence, the complaint is confirmed.It is possible that excessive force was exerted at an off angle on the device causing it to break.Other than this possibility, we cannot discern a root cause for the failure.Furthermore, no lot number was supplied which precludes conducting a device history record review or a lot specific search in the depuy synthes mitek complaints handling system.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).The lot number is not currently available.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by dfa.Depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.H10 additional narrative: if information is obtained that was not available for the initial medwatch.A follow-up medwatch will be filed as appropriate.Investigation summary the complaint device was received and evaluated.Visual observation confirms the distal tip of the driver is broken but not returned with the device.Hence, the complaint is confirmed.It is possible that excessive force was exerted at an off angle on the device causing it to break.Other than this possibility, we cannot discern a root cause for the failure.Furthermore, no lot number was supplied which precludes conducting a device history record review or a lot specific search in the depuy synthes mitek complaints handling system.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).-incomplete the lot number is not currently available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SURGICAL SCREWDRIVER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key7947289
Report Number1221934-2018-54879
Device Sequence Number2338285
Product Code LXH
UDI-Device Identifier10886705005420
UDI-Public10886705005420
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number219124
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2018
Initial Date Received by Manufacturer 09/27/2018
Supplement Date Received by Manufacturer11/07/2018
11/07/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date11/09/2018
11/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-