Model Number 10611
Medical Device Problem Codes Device Damaged Prior to Use (2284); Material Deformation (2976)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.
 
Event or Problem Description
It was reported that stent damage occurred.A 2.25 x 20 synergy stent was selected for use; however, the stent strut was noted to be broken.The device never entered the patient's body.No patient complications were reported.
 
Event or Problem Description
It was reported that stent damage occurred.A 2.25 x 20 synergy stent was selected for use; however, the stent strut was noted to be broken.The device never entered the patient's body.No patient complications were reported.Synergy ii us otw 2.25 x 20 mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found distal stent damage with stent struts lifted.The undamaged crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the shaft polymer extrusion found a shaft polymer extrusion kink 32mm distal from the distal edge of the strain relief and 971mm distal from distal edge of the strain relief.No other issues were identified during the product analysis.
 
Additional Manufacturer Narrative
Device is a combination product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key7947687
Report Number2134265-2018-61740
Device Sequence Number857356
Product Code NIQ
UDI-Device Identifier08714729840626
UDI-Public08714729840626
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/24/2020
Device Model Number10611
Device Catalogue Number10611
Device Lot Number0022121101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Initial Date Received by Manufacturer 09/14/2018
Supplement Date Received by Manufacturer10/18/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date11/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No