MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number SMXP120RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun, but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.The device history record (dhr) was reviewed, and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) caucasian female patient underwent a breast reconstruction primary with an artoura breast tissue expander 375cc and experienced rupture on the left side.Rupture was confirmed by physician during an exam.As a result, the patient underwent removal and replacement with an artoura breast tissue expander 375cc, catalog number smxp120rh, serial number (b)(4), lot number 7615733 on (b)(6) 2018.
 
Additional Manufacturer Narrative
Device evaluation summary: the device contained no fluid, upon receipt by mentor.White material was observed within the device.No foreign material was observed on the shell surface.During visual examination, mentor product analysis lab observed the device appears intact.Because the authorization for return and examination of medical device form was not signed and/or returned to mentor, mentor product analysis lab was precluded from further evaluating the device.No anomalies were found.Mentor products are 100% visually inspected prior to release in addition to thorough in-process testing during several stages of the manufacturing process.The complaint of deflation was not confirmed since no anomalies were found on the reported device.At this point it is not possible to determine the origin of the white material observed.Breast implants are not considered lifetime devices and deflation is a known complication associated with these devices and is referenced in our product insert data sheet.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key7947828
Report Number1645337-2018-06024
Device Sequence Number2338264
Product Code LCJ
UDI-Device Identifier00081317024309
UDI-Public00081317024309
Combination Product (Y/N)N
Initial Reporter StateVT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/04/2022
Device Catalogue NumberSMXP120RH
Device Lot Number7568872
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Initial Date Received by Manufacturer 09/18/2018
Supplement Date Received by Manufacturer10/16/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date11/09/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 YR
-
-