• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA HELICAL BLADE 110MM STERILE; ROD, FIXATION, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA HELICAL BLADE 110MM STERILE; ROD, FIXATION, INTRAMEDULLARY Back to Search Results
Model Number 04.038.310S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Discomfort (2330); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of implantation is an unknown date in (b)(6) 2017.Complainant part is not expected to be returned for manufacturer review/investigation.Date of concomitant therapy is the same as date of implantation, an unknown date in (b)(6) 2017.Part: 04.038.310s; lot: h261199; part manufacturing date: january 20, 2017; manufacturing site: elmira; part expiration date: december 31, 2026; nonconformance noted: n/a a review of the device history record revealed no complaint related anomalies.The device history record shows lot h261199 of tfna helical blades was processed through the normal manufacturing and inspection operations with no rework or nonconformances noted.This lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record(s) determined the raw material lot h077914 met all specifications with no issues documented that would contribute to this complaint condition.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the patient underwent hardware removal of trochanteric fixation nail advanced (tfna) implants due to discomfort in the operative hip.The helical blade implant had cut-out into the acetabulum, causing irritation.Patient was revised to total hip arthroplasty.Tfna hardware was originally implanted in (b)(6) 2017.Procedure outcome and patient status are unknown.Concomitant devices: 5.0mm titanium (ti) locking screw-sterile (part: 04.005.528s, lot: h329932, quantity: 1), 10mm/130 degree ti cannulated tfna nail (part: 04.037.042s, lot: h346500, quantity 1).This report is for a tfna helical blade.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA HELICAL BLADE 110MM STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7947857
Report Number2939274-2018-54265
Device Sequence Number19417481
Product Code HSB
UDI-Device Identifier10886982099075
UDI-Public(01)10886982099075
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K131548
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.038.310S
Device Catalogue Number04.038.310S
Device Lot NumberH261199
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2018
Initial Report FDA Received Date10/09/2018
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient Weight132
-
-