MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PROMUS ELEMENT; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Model Number 9305
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/11/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.
 
Event Description
Reportable based on device analysis completed on 11-sep-2018.A 2.50x16mm promus element stent was received with no reported issues.However, returned device analysis revealed hypotube fracture.Promus element, mr, ous 2.50x16mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.A visual and microscopic examination of the bumper tip showed no signs of damage.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple severe hypotube kinks and a partial hypotube break 585 mm distal from the distal end of the strain relief.A visual and tactile examination of the shaft polymer extrusion found no issues along the midshaft, inner or outer shaft polymer extrusion.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELEMENT
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key7947892
MDR Text Key123412220
Report Number2134265-2018-61619
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/15/2018
Device Model Number9305
Device Catalogue Number9305
Device Lot Number0019749449
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Date Manufacturer Received09/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-