Model Number N/A
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2018-05569.Concomitant medical products: psn p-ref sizer, item# 42509904000, lot# 56575104.Foreign- events occurred in: (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the posterior reference sizer¿s locking mechanism moves and unlocks when drilling the pilot holes to set the external rotation.No additional information is available at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.The reported event was considered confirmed as visual examination of the returned product identified the p-ref sizer showed nicks and gouges indicative of repeated use.No visible signs of wear were noted on the teeth.Device history record (dhr) was reviewed for deviations and/ or anomalies with no relevant deviations / anomalies identified.Per the instrument/provisional use and care package insert, instruments should be carefully inspected before each use and should not be used if the instruments are marred or worn.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PSN P-REF SIZER BOOM
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7948257
MDR Text Key123806362
Report Number0001822565-2018-05570
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
PK172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42509904010
Device Lot Number63155484
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/19/2018
Initial Date FDA Received10/09/2018
Supplement Dates Manufacturer Received07/15/2019
Supplement Dates FDA Received07/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age63 YR
-
-