| Catalog Number |
0070400 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the evacuator was difficult to suction during use.Another device was used.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed.Visual evaluation of the returned sample noted one opened (no original packaging present), bulb evacuator.Concomitant, in-house evacuator tubing was the bulb evacuator.The other end of the tubing was placed in a reservoir of water.The empty port was opened and the bulb was squeezed.The port was then closed and the bulb released to suction water.The device only suctioned 300cc of water.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿indications for use: wound drains are used to remove exudates from wound sites.A.Drain placement for 100cc, 200cc, and 400cc silicone evacuators 1.Place wound drain(s) within critical fluid collection areas.2.Draw non-perforated section of wound drain through skin until drain indicator mark appears at the skin surface.3.Attach non-perforated section of drain directly to evacuator inlet port.4.200cc and 400cc silicone evacuators: when using 2 silicone drains with one evacuator, clip sealed inlet port and attach second drain.B.Drain placement for all other evacuators 1.Place wound drain(s) within critical fluid collection areas.2.Draw non-perforated section of wound drain through skin until drain indicator mark appears at the skin surface.3.Attach non-perforated section of drain either to y-connector or directly to evacuator inlet port.4.With two silicone drains, attach blue adaptors to drains and y-connector, and attach y-connector to inlet port.Note: for 1/8¿ (3.2 mm) round drain, y-connector or enclosed 1/8¿ (3.2 mm) drain adapter must be used to connect to evacuator.Caution: do not puncture or perforate drain.C.With silicone round double drain 1.See instructions with drain.D.Attaching to auxiliary suction 1.Connect suction tube to empty port using a stepped 5-in-1 connector.2.During auxiliary suction, evacuator will deflate and exudate will flow through evacuator into suction tube.E.To establish suction 1.Open empty port.2.Squeeze evacuator.3.Close empty port.Note: reflux of fluid to the patient is minimized during reactivation by an anti-reflux valve in inlet port.F.To empty container 1.Open empty port over collection basin.2.Squeeze evacuator to empty.G.To re-establish suction 1.Repeat step ¿e¿ above.H.To read fluid volume 1.Invert unit.2.Open empty port to release vacuum.3.Read and record approximate volume.4.Empty and reactivate evacuator.".
|
| |
|
Event or Problem Description
|
|
It was reported that the evacuator was difficult to suction during use.Another device was used.
|
| |
|