| Catalog Number |
VBJR062502A |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records for the device verified the lot met all pre-release specifications.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2018, a patient was undergoing of a superficial femoral artery occlusion with a gore® viabahn® endoprosthesis.The lesion was predilated.The device was being advanced over a v18 wire through a 6fr pinnacle sheath.At some point the device separated from the delivery system.No device expansion had occurred.The device and sheath were removed as a unit.Wire access was lost.The case was completed with two new viabahn devices.The outcome was successful.
|
| |
|
Additional Manufacturer Narrative
|
|
Results code 2: 213: the engineering evaluation stated the following: the entire deice was returned.There was a kink in distal shaft, upon which the endoprosthesis is mounted, near the transition.There was approximately 1 cm of distal shaft exposed near tip.The knob doesn¿t appear to have been pulled from hub since the deployment line is taut within the hub.There were no separations in the delivery catheter.There was no deployment of the endoprosthesis noted.Engineering evaluation conclusion is inconclusive as it relates to the event description.Based on the device examination performed, no manufacturing anomalies were identified.
|
| |
|