AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550350-12
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 09/21/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported failure to advance, stent dislodgement, difficult to remove, foreign body in patient and subsequent treatment appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the heavily tortuous, heavily calcified, right coronary artery.The 3.5x12 mm xience sierra stent delivery system (sds) was advanced in the patient anatomy; however the sds failed to cross the target lesion due to resistance felt with a previously implanted stent.Resistance was felt with the previously implanted stent during removal, and the sierra stent dislodged in the implanted stent.An unspecified balloon was used to deploy the sierra stent inside the implanted stent.The lesion was treated with a 4.0mm balloon.The patient is stable.There was clinically significant delay during the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7948593
Report Number2024168-2018-07785
Device Sequence Number1500083
Product Code NIQ
UDI-Device Identifier08717648227400
UDI-Public08717648227400
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/16/2019
Device Catalogue Number1550350-12
Device Lot Number8062741
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2018
Initial Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-