• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. SENSAR; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. SENSAR; MONOFOCAL IOLS Back to Search Results
Model Number AAB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 08/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of birth: unknown, not provided.Sex/gender: unknown, not provided.If implanted, give date: not applicable as the lens was not implanted.If explanted, give date: not applicable as the lens was not implanted and therefore not explanted.(b)(6).Attempts have been made to obtain missing information; however, to date no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that during the implantation of the intraocular lens (model aab00 +19.5 diopter) breakage of the edge of the patient's anterior capsular bag was noticed.Reportedly, a 3-piece lens was implanted instead.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint could not be verified.Manufacturing record review: the manufacturing records for the product was reviewed.The product was manufactured and released according to specifications.A search revealed that no other complaints have been received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7948595
Report Number3011852734-2018-00186
Device Sequence Number857058
Product Code HQL
UDI-Device Identifier05050474548657
UDI-Public(01)05050474548657(17)200824
Combination Product (Y/N)Y
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/03/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2020
Device Model NumberAAB00
Device Catalogue NumberAAB0000195
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer01/30/2019
10/25/2020
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date02/14/2019
11/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Date Report Sent to FDA01/01/2005
-
-