CONMED CORPORATION 4.5MM STERLING LIGHTNING BUR; ARTHROSCOPIC BUR/SHAVER Back to Search Results
Catalog Number H9133
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Examination of the returned used device found the bur detached from the inner hub due to oversized inner hub dimension.A review of the device history report found no anomalies that would contribute to this issue.A 2 year review of complaint history revealed a total of 45 reports with a quantity of 52 devices related to this device family and failure mode.During the same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following; do not use equipment if upon receipt, package is opened, damaged, or shows signs of tampering.Continually check for overheating, if overheating is sensed immediately discontinue use.Do not use burs for plunge cutting.Injury or damage could occur.Do not use shaver bur if they show any signs of damage.Inspect for bent, dull, or damaged blades and burs before each use.Do not apply excessive loading/force on the shaver blade or bur.Direct contact of the rotating cutting edge of shaver blades and burs with metallic surfaces and/or other hard surfaces such as arthroscopes, cannulas, or other instruments can cause damage to the devices, shaver blade or bur should be examined for damage and replaced if necessary.A determination for further investigation has been initiated.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The conmed representative reported that his customer was using the h9133 bur during a shoulder acromioplasty and it failed within 23 seconds of use.There was no fragmenting, the bur stayed intact.There was no reported patient injury and the surgery was completed as planned.This report is being raised based on a malfunction with potential for injury upon reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM STERLING LIGHTNING BUR
Common Device Name
ARTHROSCOPIC BUR/SHAVER
Manufacturer (Section D)
CONMED CORPORATION
11311 concept boulevard
largo FL 33773
MDR Report Key7948623
Report Number1017294-2018-00157
Device Sequence Number2337924
Product Code GFF
UDI-Device Identifier10845854029671
UDI-Public(01)10845854029671(17)230517(10)932460
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2023
Device Catalogue NumberH9133
Device Lot Number932460
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2018
Initial Report FDA Received Date10/09/2018