MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC S7; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC S7; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733148
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The instrument was returned to the manufacturer for evaluation.After visual/physical examination and functional testing it was found that the instrument was slightly bent and that there was debris in the threads of the adjustment screw causing difficulties closing it.The instrument was replaced and the system with the new instrument were working as intended.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the spine clamp required replacement.There was no patient present when this issue was observed.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Updated initial reporter information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S7
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key7948649
Report Number1723170-2018-05033
Device Sequence Number2255014
Product Code HAW
UDI-Device Identifier00613994201263
UDI-Public00613994201263
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 03/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733148
Device Catalogue Number9733148
Device Lot Number121012
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer03/21/2019
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date03/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-