Device alarmed a33 during the prime test due to due to faulty force sensor resistor.Unable to perform the occlusion test, excessive no delivery test and the delivery accuracy test due to a33 alarm.All operating currents are within specification.Off no power alarm functions properly.Device passed the displacement test, rewind, basic occlusion test, self test and a21 error test.Device uploaded properly using care link.Device had minor scratched display window, a small crack on the display window, cracked case at display window corner, cracked battery tube threads, broken belt clip slot, scratched reservoir tube window, cracked reservoir tube lip and a stained end cap sticker.
|