• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN DOMINICAN REPUBLIC INC. EXTENSION SET; SET, ADMINISTRATION, INTRA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN DOMINICAN REPUBLIC INC. EXTENSION SET; SET, ADMINISTRATION, INTRA Back to Search Results
Catalog Number 490203
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: it was reported that during use of the extension set leaked blood.The leaking is occurring at the connection of the luer ending of the extension set and the patient's catheter hub.No injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Ten (10) used samples with no packaging were returned for evaluation.All ten samples were tested for leakage per specification and 3 of the 10 failed.Leakage was observed at the distal connector.The iso luer/taper testes was performed and the same 3 of 10 failed.The reported defect of leakage was confirmed through functional testing.Incidents of this nature are attributed to operator oversight during the manual assembly of the product.Although our training procedures ensure that all of our employees are properly trained in their areas of responsibility, an oversight on the part of the operator can attribute to an incident of this nature.All available information regarding this occurrence has been forwarded to our material review broad (mrb) business unit leaders and all personnel involved in the manufacturing and inspection of this product in order to heighten awareness.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If a sample and/or any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRA
Manufacturer (Section D)
B. BRAUN DOMINICAN REPUBLIC INC.
las americas industrial park
km22 autopista las americas
santo domingo,
DR 
MDR Report Key7949040
Report Number9614279-2018-00163
Device Sequence Number1343723
Product Code FPA
UDI-Device Identifier04046964303869
UDI-Public04046964303869
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K083723
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2021
Device Catalogue Number490203
Device Lot Number0061626605
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2018
Initial Date Received by Manufacturer 09/10/2018
Supplement Date Received by Manufacturer09/10/2018
Initial Report FDA Received Date10/09/2018
Supplement Report FDA Received Date11/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-