|
A manufacturing record review was completed and zero nonconformances were found, therefore supporting the device met material, assembly and performance specifications.A returned product evaluation was completed and the failure was confirmed as the distal tip of the turnpike was separated.In addition, the damage found to the turnpike shaft and distal section is consistent with torqueing the catheter against resistance.The ifu warns, "never advance, withdraw, or rotate an intravascular device against resistance until the cause of the resistance is determined by fluoroscopy.Movement of the catheter or guidewire against resistance may result in separation of the catheter or guidewire tip, other device damage, or vessel injury.".
|