• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Codes Headache (1880); Polydipsia (2604)
Date of Event 09/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(manufacturer narrative = t, corrected data = f) (b)(4).Product not returned for evaluation.No replacement product needed at this time.Customer was educated about proper technique.Most likely underlying root cause: mlc-62-user had poor technique.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure the customer's condition since the initial call and that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
 
Event or Problem Description
Consumer reported complaint for high blood glucose results.The customer is concerned with all results obtained.The expected fasting blood glucose test result range is 150mg/dl.The customer did report symptoms of headache and extreme thirst.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room/ living room.During the call on (b)(6) 2018, a blood test was performed fasting by the customer and produced test results of 183mg/dl using true metrix meter.The test strip lot manufacturer's expiration date is 07/29/2019 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Customer stated that strips are stored in the dining room/living room but was using alcohol pads when testing.Educated customer on the proper technique and hand washing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7949357
Report Number1000113657-2018-01013
Device Sequence Number13688156
Product Code NBW
UDI-Device Identifier00052569138557
UDI-Public(01)00052569138557
Combination Product (Y/N)N
Initial Reporter StateIA
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2766
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/13/2018
Initial Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-