| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Codes |
Headache (1880); Polydipsia (2604)
|
| Date of Event |
09/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not returned for evaluation.No replacement product needed at this time.Customer was educated about proper technique.Most likely underlying root cause: mlc-62-user had poor technique.Test strip udi# (b)(4).Note: manufacturer contacted customer (several attempts) in a follow-up call to ensure the customer's condition since the initial call and that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.Product notification letter sent to customer to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose results.The customer is concerned with all results obtained.The expected fasting blood glucose test result range is 150mg/dl.The customer did report symptoms of headache and extreme thirst.Medical attention is not reported as a result of the actual blood glucose results and reported symptoms.The product is stored by customer according to specification in the dining room/ living room.During the call on (b)(6) 2018, a blood test was performed fasting by the customer and produced test results of 183mg/dl using true metrix meter.The test strip lot manufacturer's expiration date is 07/29/2019 and open vial date is one month ago.The meter memory was reviewed for previous test result history: (date/time not stored correctly) (b)(6).Customer stated that strips are stored in the dining room/living room but was using alcohol pads when testing.Educated customer on the proper technique and hand washing.
|
| |
|
Search Alerts/Recalls
|