Model Number 7070M-90
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Undesired Nerve Stimulation (1980)
Date of Event 09/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmeas, there is no indication of device failure and no indication that the device was out of specification.The most probable root cause is user error; improper implant size selection, using an implant that was too long.
 
Event or Problem Description
The patient had left side si joint arthrodesis in (b)(6) 2018 where three implants were placed.The patient complained of pain following the initial procedure.The surgeon determined that the most cranial positioned implant was too long and was impinging on the neuroforamen.Four days after the initial procedure, the surgeon performed a revision procedure where he removed the cranial positioned implant and replaced it with a shorter implant of the same type in a new position.No other preexisting implants were adjusted or removed.The patient's symptoms resolved following the revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IFUSE IMPLANT SYSTEM
Common Device Name
ORTHOPEDIC ROD
Manufacturer (Section D)
SI-BONE, INC.
471 el camino real
suite 101
santa clara CA 95050 4482
MDR Report Key7949513
Report Number3007700286-2018-00087
Device Sequence Number1499723
Product Code OUR
UDI-Device Identifier00851085007459
UDI-Public00851085007459
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K080398
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/13/2022
Device Model Number7070M-90
Device Lot Number2615801
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2018
Initial Report FDA Received Date10/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-