| Model Number |
15 |
| Medical Device Problem Code |
Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer¿s device and verified the reported issue.The device had a large amount of water inside and further troubleshooting was not possible.The customer has taken the device out of service.The cause of the reported issue could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device displayed a service wrench.Upon evaluation of the customer¿s device, physio-control observed that the device displayed an "abnormal energy delivered" message.As a result, the wrong defibrillation or no defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
|
| |
|