| Model Number |
AAB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
07/25/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If implanted, give date: not applicable as the lens was not implanted.If explanted, give date: not applicable as the lens was not implanted and therefore not explanted.(b)(6).Attempts have been made to obtain missing information; however, to date no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during surgery, the patient's posterior capsule was ruptured.Reportedly, the intraocular lens (model aab00 +19.5 diopter) was not implanted.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|